T-LDRS4810-H2O
Ref:T-LDRS4810_H2O_IntE
Quoted Date: January 24, 2006
__for Moisture on-line analysis of solids applications
![]()
l Intelligent Analysis Based on Powerful CPU, Reliable, solid state design
l Rapid Test, Lab Accurate
l Proven technology, Multi-parameters Tested for Compensation
l 100 frame Data log
l LCD 2x16 Display, Backlight Available
l RS232/485 Communication, printer supported by RS232C
l Full Function Keypad Operation
l Memo Driven Software Operation and Setup
![]()
The LDRP Reflectance Spectrometer is a portable instrument designed for industry field and laboratory investigations of chemical composition both of liquids and solids materials.
Technically, the LDRP is a multi-band photometer rather than a spectrometer, because it does not measure reflectance across a continuum of wavelengths. Rather, it measures reflectance in separate, non-overlapping ranges of wavelength. However, this distinction is not important for most applications.
The LDRP consists of lamps, a photo-detector, and an analyzer with display. These elements are arranged and coordinated with the sample and with each other to provide reproducible, easily interpretable results.
The Lamps are arranged so that their light passes are the same for each wavelength, being directed on to the surface of samples to be measured. And the diffused light after absorption by the samples was collected with focus system to the sensor.
In this configuration, the LDRP works best and most reproducibly with flat objects that have matte surfaces. Some shininess (as in most leaves) is acceptable.
Mirror-like surfaces will yield spurious readings.
For moderately rough surfaces, as small particles, the LDRP yields reasonable values for reflectance.
With very rough samples, light leaks can be a problem. Users can use the light protector sleeve to stop light leaks.
BD5-LDRP481X is special to samples. Users always need calibration for special applications. Internal memory capable of storing 8 separate calibrations. Use the calibrations provided by system, or create your own calibrations using the software function provided.
LDRP481X is designed for absorption bands detection, LDRS482X is designed for specific molecular absorption line analysis.
|
|
|
l Instant Measurement
l Non-Contact
l Non-Destructive
l Versatile usage
Quantitative moisture analysis for solids and liquids, with high selectivity.
The high OH absorption coefficient in the NIR Region makes BDTI Vis/NIR analyzers an ideal tool for accurate and flexible moisture analysis in raw materials, in-process and finished product. Using NIR as opposed to traditional, lengthy, and costly traditional moisture analysis, could save significant money and time.
Nutriment With growing pressure from the Food and Drug Administration on nutriments companies to standardize ingredients and ensure product purity, NIR technology has emerged as a cost-effective means for quality control in the manufacturing process of natural supplements. With the wider spectral range of LDRP and LDRS that other NIR instruments available, BDTI is taking 100% inspection of raw materials and finished products to a higher level by providing a vehicle for more reliable spectral data collection and seamless user-friendly interfaces to available chemo metric software. Professional food Nutrimental Analysis, enabling raw material testing quickly—at the dock, on the manufacturing floor, or where ever it is needed—positioning you to meet GMP and FDA requirements, quickly and easily.
Examples:
l Chemicals, Ink
l Pulp/Paper
l Pharmaceuticals
l Food, Beverages and Dairy
l Agriculture Grain, soil, woods, fruits, Greenstuffs
l Cement and other materials
l Cosmetics and Beauty Products
l Lubricant Oils, machine oils, fuel oils
l Plastics, polymers
Moisture Range:
Solids: 0.01-100%;
Liquid Range: 0.01-100/1000ppm; to 100%
Accuracy: <±1%Rel or 0.01% abs, whichever great;
Outputs: 2x16 LCD display,
Data Transfer: Via RS-485 port for interface with PC
Calibrations Stores up to 8 calibration sets
Data Storage: Stores data for up to 100 samples
Optical Capabilities
Measurement Mode: Diffuse transmittance, Diffuse reflectance
Scan Speed: 100ms;
Optical Range:0 to 6 AU
Resolution:0.00001 AU
Stability:0.05 Mili-AU
Measurement Time: Variable (Typically under 20 seconds)
Installation
Position: above samples and/or flow;
Fittings: bolts to suitable frame;
Sample Information
Thickness >75 mm
Distance:300-400mm (from sensor to the sample surface;
Scanning area: φ50mm;
Sample size:: unlimited( but different samples must be calibrated).
Pre process: None;
Sensor Environ Temperature: 0-40℃;
Sensor Environ Humidity: 0-95%RH(none condensing, usually dry air blowing system needed);
Physical Data
Dimensions:
Analyzer: 125mm X 180mm X 60mm
Sensor: Æ56x350 mm
Weight: less than 7.5 kg
Supply:
Inside Power Supply: 24V DC;
AC Power Requirements....................................120 or 240 VAC 50/60 Hz AC adapter (supplied)
DC Power Requirements ................................... (included)
Power Consumption .......................................... 24V, MAX<10W;
Package: IP56
BigDipper Technochem Institute
Call:86-10-8264.0226; Fax: 86-10-8264.0221;
P.o.Box 603 BDTI Beijing, China 100080
email: sun@bigdipper-technochem.com suncns@hotmail.com suncns@yahoo.com